Operating the instrument
R&S
®
FE50DTR
28
Manual 1179.3180.02 ─ 02
Table 6-1: Signals and connections in Rx mode
Signal
R&S
FE50DTR
Vector signal analyzer
e.g, R&S
FSV/A
Reference
"Ref. In"
Connector "REF Output"
Freq.: 640
MHz, 10
MHz
LAN (Control)
"LAN"
"LAN"
IF Out
"IF Out A"
"RF Input 50
Ω
"
RF In
"RF A"
-
How to:
●
Chapter 4.7, "Connecting to LAN"
●
Chapter 4.8, "Connecting to SMA and 1.85
To control the R&S
FE50DTR with a vector signal analyzer
The R&S
FSV/A controls the R&S
FE50DTR using a secure socket layer (SSL)
control connection via LAN. This step-by-step description explains how the
R&S
FSV/A locks the R&S
FE50DTR to control the R&S
FE50DTR exclusively.
1. Check, that no other instrument controls the R&S
FE50DTR.
The LAN LED on the front panel of the R&S
FE50DTR must be green, see
2. At the R&S
FSV/A, open a measurement channel for an I/Q-based applica-
tion, such as the I/Q Analyzer ([Mode]).
3. Select [Input / Output] > "Input Source Config" > "Input Source" > "External
Frontend".
4. On the "Global Config" tab, enable general use of the external frontend for the
application ("On").
The connection settings become available.
5. Select the "Type" of connected frontend.
6. Enter the IP address or computer name (hostname) of the connected
R&S
FE50DTR.
You can find the information on the electronic label on the side panel of the
R&S
Chapter 5.3, "Electronic label"
Operating the R&S
FE50DTR in Rx mode