Configuring the Input Signal and Measurement Mode
R&S
®
EVSF1000
33
User Manual 1178.6410.02 ─ 05
5
Configuring the Input Signal and Measure-
ment Mode
The following settings are common to all measurement modes. They define which data
is measured and how, and are indicated at the top of the measurement window.
............................................................................................... 33
................................................................................................ 34
5.1
Input Signal Settings
Access VNC
: [n] > [F1]
The following settings configure the receiver signal to be measured by the
R&S
EVSF1000.
Correction RX 1 In
Configures the correction of the RX 1 input signal (in dB).
This value is subtracted from all levels and compensates effects of upstream lines and
attenuators.
Correction RX 2 In
Configures the correction of the RX 2 input signal (in dB).
This value is subtracted from all levels and compensates effects of upstream lines and
attenuators.
5.2
Receiver Board
If a second receiver board is installed (requires the R&S
EVSG-B1 option), each
receiver board can be configured to perform different measurement tasks.
Receiver Board