Remote Control
R&S
®
ESRP
811
User Manual 1175.7445.02 ─ 08
[SENSe:]SWEep:EGATe:TYPE
<Type>
This command sets the type of triggering by the external gate signal.
A delay between applying the gate signal and the start of recording measured values
can be defined, see
Parameters:
<Type>
LEVel | EDGE
LEVel
The gate is level-triggered:
After detection of the gate signal, the gate remains open until the
gate signal disappears. The gate opening time cannot be
defined with the command
.
Note
: Using gating with gate mode "level" and an IFP trigger
holdoff time for the IFP trigger is ignored for frequency sweep,
FFT sweep, zero span and IQ mode measurements.
EDGE
The gate is edge-triggered:
After detection of the set gate signal edge, the gate remains
open until the gate delay (
has expired.
*RST:
EDGE
Example:
SWE:EGAT:TYPE EDGE
Manual operation:
See
11.5.2.6
Vertical Axis Configuration
The following commands configure the characteristics of the vertical diagram axis.
Commands useful to configure the vertical axis described elsewhere:
●
DISPlay[:WINDow<n>]:TRACe<t>:Y:SPACing
●
●
●
●
........................................................................................... 812
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
................................................................... 812
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:MODE
.........................................................812
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel
........................................................813
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet
........................................... 813
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
................................................... 814
............................................................................................... 814
Remote Commands in Spectrum Analyzer Mode