Remote Control
R&S
®
ESRP
804
User Manual 1175.7445.02 ─ 08
Example:
SWE:TYPE FFT
Selects FFT sweeps.
Manual operation:
See
See
See
[SENSe:]SWEep:TYPE:USED?
This command queries the sweep type if you have turned on automatic selection of the
sweep type.
Return values:
<Type>
SWE | FFT
SWE
Normal sweep
FFT
FFT mode
*RST:
SWE
Example:
SWE:TYPE:USED?
Usage:
Query only
11.5.2.5
Trigger Configuration
The following commands configure triggered or gated measurements.
Triggering Measurements
TRIGger<n>[:SEQuence]:HOLDoff[:TIME]
........................................................................ 804
TRIGger<n>[:SEQuence]:IFPower:HOLDoff
..................................................................... 805
TRIGger<n>[:SEQuence]:IFPower:HYSTeresis
................................................................. 805
TRIGger<n>[:SEQuence]:LEVel[:EXTernal]
.......................................................................805
TRIGger<n>[:SEQuence]:LEVel:IFPower
..........................................................................806
TRIGger<n>[:SEQuence]:LEVel:RFPower
........................................................................ 806
TRIGger<n>[:SEQuence]:LEVel:VIDeo
.............................................................................806
.....................................................................................807
.................................................................................. 807
TRIGger<n>[:SEQuence]:TIME:RINTerval
........................................................................ 808
TRIGger<n>[:SEQuence]:HOLDoff[:TIME]
<Delay>
This command defines the length of the trigger delay.
A negative delay time (pretrigger) can be set in zero span only.
Suffix:
<n>
.
irrelevant
Remote Commands in Spectrum Analyzer Mode