Remote Control
R&S
®
ESRP
803
User Manual 1175.7445.02 ─ 08
[SENSe:]SWEep:TIME
<Time>
This command defines the sweep time.
The range depends on the frequency span.
If this command is used in "Spectrum" mode, automatic coupling to resolution band-
width and video bandwidth is switched off.
Parameters:
<Time>
refer to data sheet
*RST:
(automatic)
Example:
SWE:TIME 1s
Manual operation:
See
See
See
[SENSe:]SWEep:TIME:AUTO
<State>
In realtime mode, this command automatically sets the sweep time to 32
ms.
In analyzer mode, this command controls the automatic coupling of the sweep time to
the frequency span and bandwidth settings. If
is used, auto-
matic coupling is switched off.
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
SWE:TIME:AUTO ON
Activates automatic sweep time.
Manual operation:
See
See
See
[SENSe:]SWEep:TYPE
<Type>
This command selects the sweep type.
Parameters:
<Type>
SWE
Selects analog frequency sweeps.
AUTO
Automatically selects the sweep type (FFT or analog frequency
sweep).
FFT
Selects FFT sweeps.
*RST:
AUTO
Remote Commands in Spectrum Analyzer Mode