Remote Control
R&S
®
ESRP
733
User Manual 1175.7445.02 ─ 08
Manual operation:
See
[SENSe:]ESPectrum:RANGe<range>[:FREQuency]:STARt
<Frequency>
This command sets the start frequency for the specified range.
In order to change the start/stop frequency of the first/last range, select the appropriate
span. If you set a span that is smaller than the overall span of the ranges, the mea-
surement includes only the ranges that lie within the defined span and have a minimum
span of 20 Hz. The first and last range are adapted to the given span as long as the
minimum span of 20 Hz is not violated.
Note the rules for the <Frequency> parameter specified in
Suffix:
<range>
.
1...20
range
Parameters:
<Frequency>
numeric value
*RST:
-250.0 MHz (range 1), -2.52 MHz (range 2), 2.52
MHz (range 3)
Example:
ESP:RANG1:STAR 100000000
Sets the start frequency for range 1 to 100 MHz.
Manual operation:
See
[SENSe:]ESPectrum:RANGe<range>[:FREQuency]:STOP
<Frequency>
This command sets the stop frequency for the specified range. For further details refer
to the
[SENSe:]ESPectrum:RANGe<range>[:FREQuency]:STARt
Note the rules for the <Frequency> parameter specified in
Suffix:
<range>
.
1...20
range
Parameters:
<Frequency>
numeric value
*RST:
-2.52 MHz (range 1), 2.52 MHz (range 2), 250.0
MHz (range 3)
Example:
ESP:RANG3:STOP 10000000
Sets the stop frequency for range 2 to 10 MHz.
Manual operation:
See