[SENSe:]SCAN<range>:INPut:GAIN[:STATE]
<State>
This command turns the preamplifier in a particular scan range on and off.
Suffix:
<range>
.
1...10
Selects the scan range.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
SCAN4:INP:GAIN:STAT ON
Turns on the preamplifier for the 4th scan range.
[SENSe:]SCAN<range>:INPut:TYPE
<Input>
This command selects the signal source for a particular scan range.
Suffix:
<range>
.
1...10
Selects the scan range.
Parameters:
<Input>
INPUT1
Selects RF input 1.
INPUT2
Selects RF input 2.
*RST:
INPUT1
Example:
SCAN4:INP:TYPE INPUT2
Selects RF input 2 as the signal source for the 4th scan range.
[SENSe:]SCAN:RANGes[:COUNt]
<Ranges>
This command defines the number of scan ranges.
Parameters:
<Ranges>
Number of ranges in the scan table.
If you enter the value "0", the R&S
ESRP ignores the configura-
tion of the scan table. instead, it performs the measurement
based on the current receiver configuration.
Range:
0 to 10
*RST:
0
Example:
SCAN:RANG:COUN 4
Defines 4 scan ranges.
[SENSe:]SCAN<range>:STARt
<Frequency>
This command defines the start frequency of a particular scan range.
Remote Commands in Receiver Mode