Spectrum Measurements
R&S
®
ESRP
383
User Manual 1175.7445.02 ─ 08
●
●
●
●
●
Spectrogram:
Chapter 5.4.2.3, "Markers in Spectrograms"
Functions to position markers exclusive in Spectrum mode
.......................................................................................383
................................................................................... 383
Center =Mkr Freq (span > 0)
Sets the center frequency to the current marker or delta marker frequency. A signal
can thus be set to as center frequency, for example to examine it in detail with a
smaller span.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:CENTer
Ref Lvl =Mkr Lvl
Sets the reference level to the current marker level.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
Auto Max Peak/Auto Min Peak
Adds an automatic peak search action for marker 1 at the end of each particular
sweep. This function may be used during adjustments of a device under test to keep
track of the current peak marker position and level.
The current marker search limit settings (
taken into account.
Remote command:
CALCulate<n>:MARKer<m>:MAXimum:AUTO
CALCulate<n>:MARKer<m>:MINimum:AUTO
Exclude LO
Switches the frequency range limit for the marker search functions on or off.
"ON"
The minimum frequency included in the peak search range is ≥ 5 ×
resolution bandwidth (RBW).
Due to the interference by the first local oscillator to the first inter-
mediate frequency at the input mixer, the LO is represented as a sig-
nal at 0 Hz. To avoid the peak marker jumping to the LO signal at 0
Hz, this frequency is excluded from the peak search.
"OFF"
No restriction to the search range. The frequency 0 Hz is included in
the marker search functions.
Remote command:
CALCulate<n>:MARKer<m>:LOEXclude
Analysis