Spectrum Measurements
R&S
®
ESRP
314
User Manual 1175.7445.02 ─ 08
6.2.1.2
Initial Configuration
The initial configuration is selected such that the RF input is always protected against
overload, provided that the applied signal levels are in the allowed range for the instru-
ment.
The parameter set of the initial configuration can be customized using the "Startup
Recall" softkey in the "Save/Rcl" menu. For further information refer to
"Saving and Recalling Settings Files – SAVE/RCL Key"
Table 6-5: Initial configuration
Parameter
Setting
mode
Spectrum
sweep mode
auto
center frequency
f
max
/2
center frequency step size
0.1 * span
span
maximum possible (depends on instrument model)
RF attenuation
0 dB
reference level
-10 dBm
level range
100 dB log
level unit
dBm
sweep time
auto
resolution bandwidth
auto (3 MHz)
video bandwidth
auto (3 MHz)
FFT filters
off
span/RBW
100
RBW/VBW
1
sweep
cont
trigger
free run
trace 1
clr write
trace 2/3/4/5/6
blank
detector
auto peak
frequency offset
0 Hz
reference level offset
0 dB
reference level position
100 %
grid
abs
cal correction
on