![Rohde & Schwarz CMW500 Series User Manual Download Page 245](http://html1.mh-extra.com/html/rohde-and-schwarz/cmw500-series/cmw500-series_user-manual_1477718245.webp)
Command Reference
R&S
®
CMW500
245
User Manual 1173.9463.02 ─ 02
Example:
SYSTem:COMMunicate:NET:HOSTname
'hh303132.company.net'
Define the host name.
Firmware/Software:
V1.0.0.4
SYSTem:COMMunicate:NET:DHCP
<Boolean>
Enables or disables the Dynamic Host Configuration Protocol (DHCP).
Parameters:
<Boolean>
-1 | 0
-1:
DHCP enabled, automatic TCP/IP address setting
0:
DHCP disabled, manual address setting
*RST:
a *RST does not affect the LAN service settings
Example:
SYSTem:COMMunicate:NET:DHCP 0
Disable DHCP, enable manual setting of the IP address informa-
tion
Firmware/Software:
V1.0.0.4
SYSTem:COMMunicate:NET:IPADdress
'<Address>'
Defines the IP address of the R&S CMW500. This address is valid if DHCP is disabled
(
).
Parameters:
'<Address>'
String parameter, IP address consisting of four blocks (octets)
separated by dots.
*RST:
a *RST does not affect the LAN service settings
Example:
SYSTem:COMMunicate:NET:IPADdress '10.113.10.38'
Select a private IP address (characterized by 10 in the first octet).
Firmware/Software:
V1.0.0.4
SYSTem:COMMunicate:NET:SUBNet:MASK
'<Address>'
Defines the subnet mask address of the R&S CMW500. This address is valid if DHCP is
disabled (
).
Parameters:
'<Address>'
String parameter, subnet mask consisting of four blocks separated
by dots.
*RST:
a *RST does not affect the LAN service settings
Example:
SYSTem:COMMunicate:NET:SUBNet:MASK
'255.255.0.0'
Select a subnet mask. With this subnet mask, the first two octets
Firmware/Software:
V1.0.0.4
Instrument-Control Commands