Command Reference
R&S
®
CMW-KM750/-KM751
86
User Manual 1173.9670.02 ─ 05
Parameters:
<HighDynAsemLev> Range:
-60 dB to -10 dB
*RST:
-60 dB, OFF
Default unit: dB
Additional parameters: OFF | ON (disables | enables the limit
check using the previous/default limit values)
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V1.0.4.11
Manual operation:
See
"Power vs Time > Two Shot Assembly Level"
CONFigure:TDSCdma:MEAS<i>:MEValuation:SCOunt:SPECtrum
<StatCntSpec>
Specifies the statistic count of the measurement. The statistic count is equal to the num-
ber of measurement intervals per single shot. Use
CONFigure:...:MEAS<i>:...:REPetition SINGleshot | CONTinuous
to
select either single-shot or continuous measurements.
See also: "Statistical Settings" in the R&S CMW user manual, chapter "Remote Control"
Parameters:
<StatCntSpec>
Number of measurement intervals.
Range:
1 to 1000
*RST:
10
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V1.0.2.7
Manual operation:
See
"Power vs Time / Modulation / Spectrum / BER > Statistic
CONFigure:TDSCdma:MEAS<i>:MEValuation:SCOunt:BER
<StatCntBER>
Specifies the statistic count of the measurement. The statistic count is equal to the num-
ber of measurement intervals per single shot. Use
CONFigure:...:MEAS<i>:...:REPetition SINGleshot | CONTinuous
to
select either single-shot or continuous measurements.
See also: "Statistical Settings" in the R&S CMW user manual, chapter "Remote Control"
Parameters:
<StatCntBER>
Number of measurement intervals.
Range:
1 to 1000
*RST:
100
Example:
See
Specifying Additional Measurement-Specific Settings
Firmware/Software:
V1.0.5.3
Manual operation:
See
"Power vs Time / Modulation / Spectrum / BER > Statistic
Multi Evaluation Measurement Commands