Command Reference
R&S
®
CMW-KM750/-KM751
129
User Manual 1173.9670.02 ─ 05
<10_MarginAB>
<11_MarginCD>
<12_MarginDE>
<13_MarginED>
<14_MarginDC>
<15_MarginBA>
Limit line margin values in the 6 emission mask areas
Range:
-100 dB to 100 dB
Default unit: dB
Example:
See
Usage:
Query only
Firmware/Software:
V3.0.10
V3.0.30: maximum number of segments increased to 500
Options:
R&S CMW-KM012
6.3.21 List Mode Results (All Segments, One Result)
Each of the following commands returns a selected list mode result for all measured
segments. The number of returned results depends on the number of measured seg-
ments, see
To configure the list mode use the commands described in
For a description of the list mode see
chapter 3.3, "Multi Evaluation List Mode"
Indicated ranges apply to all statistical results except standard deviation results. The
minimum for standard deviation results equals 0. The maximum equals the width of the
indicated range divided by two. Exceptions are explicitly stated.
FETCh:TDSCdma:MEAS<i>:MEValuation:LIST:SRELiability?
Returns the segment reliability for all measured list mode segments.
A common reliability indicator of zero indicates that the results in all measured segments
are valid. A non-zero value indicates that an error occurred in at least one of the measured
segments. If you get a non-zero common reliability indicator, you can use this command
to retrieve the individual reliability values of all measured segments for further analysis.
Return values:
<Reliability>
<SegReliablity>
Comma separated list of values, one per measured segment
The meaning of the returned values is the same as for the common
reliability indicator, see previous parameter.
Example:
See
Usage:
Query only
Firmware/Software:
V3.0.11
Options:
R&S CMW-KM012
Multi Evaluation Measurement Commands