![Rohde & Schwarz CMW-KE100 User Manual Download Page 245](http://html.mh-extra.com/html/rohde-and-schwarz/cmw-ke100/cmw-ke100_user-manual_1477807245.webp)
1xEV-DO Measurements
R&S
®
CMW-KM8xx/-KS8xx
245
User Manual 1173.9511.02 ─ 12
As some code channels are not continuously present in the signal, the user may spec-
ify a code channel filter, see
Multi Evaluation: Code Channel Filter
measurement depending on the presence of a certain code channel (DRC, Data, ACK,
DSC, Aux Pilot) and the first / second half of a timeslot.
3.2.7 Limit Settings and Conformance Requirements
Conformance requirements for 1xEV-DO TX tests are specified in 3GPP2 C.S0033,
section 4, "Physical Layer Transmitter Minimum Standards".
The following sections give an overview of the R&S
CMW limit settings and the related
test requirements.
................................................................................... 245
..............................................................................................246
............................................................................ 247
...................................................................................247
3.2.7.1
Transmit Modulation Limits
A poor modulation accuracy of the AT transmitter increases the transmission errors in
the reverse channel of the 1xEV-DO network. The waveform quality is the critical quan-
tity to assess the modulation accuracy of a 1xEV-DO AT.
3GPP2 C.S0033-B section 4.2.2 defines the minimum requirements for the waveform
quality in terms of a scalar "waveform quality factor"
●
the waveform quality factor shall be greater than 0.944
●
the transmit time error shall not exceed 1 μs
●
the frequency error shall not exceed 300 Hz for band class groups 450 and 800
and shall not exceed 150 Hz for band class group 1900 (see
These limits can be set in the configuration dialog (the standard values being the pre-
sets) along with limits for other modulation characteristics. The checkboxes activate/
deactivate the limit checks during measurements.
General Description
深圳德标仪器