1xEV-DO Signaling
R&S
®
CMW-KM8xx/-KS8xx
105
User Manual 1173.9511.02 ─ 12
The conformance test specifications 3GPP2 C.S0032-B and 3GPP2 C.S0033-B define
a variety of test cases related to PER limits. Adjust the value in accordance with your
measurement.
Fig. 2-55: Limit settings
Multi Carrier Limit Evaluation
...................................................................................... 105
Multi Carrier Limit Evaluation
For multi-carrier measurements (Physical Layer Subtype 3) , limits can either be
applied to the individual carriers or over all carriers. Select the appropriate strategy.
Remote command:
CONFigure:EVDO:SIGN<i>:RXQuality:LIMit:PER:EVALuation
Max PER
Maximum packet error rate allowed. An error (limit failure) is indicated if the "Max PER"
is exceeded after the statistics cycle is completed (see
on page 102). In a confidence test, the AT may pass the limit check before the statis-
tics cycle is completed.
Different thresholds can be configured for FL and RL. Setting the value to 100% disa-
bles the limit check including the confidence test.
Remote command:
CONFigure:EVDO:SIGN<i>:RXQuality:LIMit:FLPer:MPER
CONFigure:EVDO:SIGN<i>:RXQuality:LIMit:RLPer:MPER
Min Confidence Level
Minimum required statistical probability for the PER to be below the "Max PER".
Different thresholds can be configured for FL and RL PER. Setting the value to 100%
disables the confidence test.
Remote command:
Suppress "Cell Off" Message
If the "1xEV-DO Signaling" softkey is selected and the cell signal is on and you press
ON | OFF, a warning can be displayed, asking whether you really want to switch off the
cell signal. In this situation, users often want to start / stop the measurement, not the
cell signal, but forget to select the "RX Meas..." softkey.
GUI Reference
深圳德标仪器