Command Reference
R&S
®
CMW-B400B/-B405A
87
User Manual 1176.8116.02 ─ 04
<Level>
Level of the tone.
For method SEP, the level can be set and queried.
For method TLEV, the level of a setting command is ignored and
a query returns the level resulting from the total multitone signal
level.
For method selection see
The total level of the multitone signal must not exceed 5 V.
Range:
0 V to 5 V
*RST:
0.05 V
Default unit: V
Example:
See
Firmware/Software:
V3.2.10
Manual operation:
See
SOURce:AUDio:GEN<i>:DIGital:MULTitone:TONE<no>:TDEFinition
<Frequency>,
<Level>
Configures a selected tone of the multitone list of the digital generator.
Suffix:
<no>
.
1..20
Selects the tone
Parameters:
<Frequency>
Frequency of the tone. All tones must have different frequencies.
Range:
0 Hz to 21E+3 Hz
*RST:
1000 * <no>
Default unit: Hz
<Level>
Level of the tone.
For method SEP, the level can be set and queried.
For method TLEV, the level of a setting command is ignored and
a query returns the level resulting from the total multitone signal
level.
For method selection see
The total level of the multitone signal must not exceed 1 FS.
Range:
0 FS to 1 FS
*RST:
0.05 FS
Default unit: FS
Example:
See
Firmware/Software:
V3.2.10
Manual operation:
See
Audio Generators