Command Reference
R&S
®
CMW-B400B/-B405A
79
User Manual 1176.8116.02 ─ 04
Parameters:
<Frequency>
OFF | LP3 | LP4 | LP15
OFF:
lowpass filter disabled
LP3 | LP4 | LP15:
cutoff frequency 3 kHz, 4 kHz or 15 kHz
*RST:
OFF
Example:
See
Firmware/Software:
V3.2.10
Manual operation:
See
CONFigure:AUDio:MEAS<i>:ANALog:FFTNoise:FILTer:HPASs
<Frequency>
CONFigure:AUDio:MEAS<i>:DIGital:FFTNoise:FILTer:HPASs
<Frequency>
Configures the cutoff frequency of the highpass filter for FFT noise measurements.
Parameters:
<Frequency>
OFF | H6 | H50 | H300
OFF:
highpass filter disabled
H6 | H50 | H300:
cutoff frequency 6 Hz, 50 Hz or 300 Hz
*RST:
ANALog = OFF, DIGital = H6
Example:
See
Firmware/Software:
V3.2.10
Manual operation:
See
CONFigure:AUDio:MEAS<i>:ANALog:FFTNoise:FILTer:WEIGhting
<Weighting>
CONFigure:AUDio:MEAS<i>:DIGital:FFTNoise:FILTer:WEIGhting
<Weighting>
Configures the weighting filter for FFT noise measurements.
Parameters:
<Weighting>
OFF | AWEight | CCITt | CMSG
OFF:
weighting filter disabled
AWEight:
A-weighting filter
CCITt:
CCITT weighting filter
CMSG:
C-message weighting filter
*RST:
OFF
Example:
See
Firmware/Software:
V3.2.10
Manual operation:
See
6.3.8 FFT Noise Measurement Results
The following commands return the results of FFT noise measurements.
Audio Measurements