ABORt:AUDio:MEAS<i>:ANALog
INITiate:AUDio:MEAS<i>:DIGital
STOP:AUDio:MEAS<i>:DIGital
ABORt:AUDio:MEAS<i>:DIGital
Starts, stops, or aborts an analog or digital audio measurement:
●
INITiate...
starts or restarts the measurement; the R&S
CMW enters the
"RUN" state.
●
STOP...
causes a running measurement to stop after the current evaluation
period is terminated and valid results are available; the R&S
CMW enters the
"RDY" state.
●
ABORt...
causes a running measurement to stop immediately; the R&S
CMW
enters the "OFF" state.
Use
FETCh...STATe?
to query the current measurement state.
See also: "Measurement Control" in the R&S
CMW user manual, chapter "Remote
Control"
Example:
See
Usage:
Event
Firmware/Software:
V3.2.10
Manual operation:
See
FETCh:AUDio:MEAS<i>:ANALog:STATe?
FETCh:AUDio:MEAS<i>:DIGital:STATe?
Queries the main measurement state. Use
FETCh:...:STATe:ALL?
to query the
measurement state including the substates. Use
INITiate..., STOP...,
ABORt...
to change the measurement state.
See also: "Measurement Control" in the R&S
CMW user manual, chapter "Remote
Control"
Return values:
<MeasState>
OFF | RUN | RDY
OFF:
measurement switched off, no resources allocated, no
results available (when entered after
ABORt...
)
RUN:
measurement running (after
INITiate..., READ...
),
synchronization pending or adjusted, resources active or queued
RDY:
measurement has been terminated, valid results may be
available
*RST:
OFF
Example:
See
Usage:
Query only
Firmware/Software:
V3.2.10