Remote Control Commands
R&S
®
AREG100A
241
User Manual 1178.7417.02 ─ 04
:SYSTem:REBoot
Reboots the instrument including the operating system.
Usage:
Event
:SYSTem:RESTart
Restarts the instrument without restarting the operating system.
Usage:
Event
:SYSTem:SHUTdown
Shuts down the instrument.
Usage:
Event
Manual operation:
See
:SYSTem:WAIT
<TimeMs>
Delays the execution of the subsequent remote command by the specified time.
This function is useful, for example to execute an SCPI sequence automatically but
with a defined time delay between some commands.
See
Chapter 7.2.3, "Assigning Actions to the [User] Key"
Setting parameters:
<TimeMs>
integer
Wait time in ms
Range:
0 to 10000
*RST:
0
Example:
:SYSTem:WAIT 10000
// waits 10s before resetting the instrument
*RST
Usage:
Setting only
9.11
TEST
................................................................................................241
:TEST<hw>:ALL:RESult?
Queries the result of the performed selftest. Start the selftest with
.
TEST