R&S
®
ENV216
Two Line V-NetworkUser Manual
5201669314Version 05
(Â1Ðë>)
Page 1: ...R S ENV216 Two Line V Network User Manual 5201669314 Version 05 1...
Page 2: ...orfstr 15 81671 Muenchen Germany Phone 49 89 41 29 0 Email info rohde schwarz com Internet www rohde schwarz com Subject to change data without tolerance limits is not binding R S is a registered trad...
Page 3: ...city can cause Follow the safety instructions pro vided in Chapter 1 1 Safety instructions on page 3 and the additional information provided during setup or operation procedures Where do I find safety...
Page 4: ...ervice personnel authorized by Rohde Schwarz are allowed to repair the LISN Contact Rohde Schwarz customer service at http www rohde schwarz com support Lifting and carrying the product Look up the we...
Page 5: ...g to power The product is an overvoltage category II product and has to be connected to a fixed installation used to supply energy consuming equipment such as household applian ces and similar loads E...
Page 6: ...e with DIN EN IEC UL CSA 61010 1 DIN EN IEC UL CSA 60950 1 or DIN EN IEC UL CSA 62368 1 Using accessories Select accessories that are suitable for the product and the measurement task espe cially if t...
Page 7: ...nding terminal earth ground contact Chassis grounding terminal 1 3 Warning messages in the documentation A warning message points out a risk or danger that you need to be aware of The sig nal word ind...
Page 8: ...Safety and regulatory information R S ENV216 8 User Manual 5201 6693 14 05 1 4 Korea certification class B B Korea certification class B...
Page 9: ...Isolation of the test circuit from interference of the power source The R S ENV216 Two Line V Network is constructed using air core inductances and conforms with the recommendations of CISPR 16 1 2 EN...
Page 10: ...device is damaged you should notify the responsible freight carrier immediately and retain all items of packaging in order to support your complaint The original pack aging is also useful for use when...
Page 11: ...1 Socket for connecting the equipment under test For country specific sockets see Chapter A Country specific Connector Models on page 30 Reference ground 2 4 mm jacks for connecting to the equipment...
Page 12: ...nsient limiter attached before the RF Output output to protect measurement equipment from high positive and negative voltage spikes 6 V By turning off the protection diodes higher interference voltage...
Page 13: ...aded bolt with nut is used only for protective grounding It is not suitable for use as a reference ground Mains connection cable 3 The mains connection cable is primarily used to supply power to the e...
Page 14: ...lug in power supply The control circuitry of the R S ENV216 is powered via the External Power Supply socket if the R S ENV216 is to be used to conduct measurements of AC voltages below 90 VAC or on DC...
Page 15: ...to the power source you have to connect the R S ENV216 to a protective ground terminal An unearthed R S ENV216 is live Touching a live electrical device causes serious personal injury or even death T...
Page 16: ...onductor that is compliant with VDE0100 Users must be aware that it is possible for socket connections and grounding conduc tors to become disconnected A further grounding conductor with adequate diam...
Page 17: ...until you have disconnected the V network from the power supply Connect the R S ENV216 to the mains power supply 2 If the phase connection is made correctly LED L1 1 must light up If it does not For m...
Page 18: ...y damage the equipment 2 DANGER Risk of electric shock Make sure that the external power supply com plies with the requirements for reinforced double insulation in accordance with DIN EN IEC UL CSA 61...
Page 19: ...ference mass are decoupled by a PE choke This choke should not be bridged by the EMI test receiver which is also connected to the measurement reference ground Therefore either an isolated or a battery...
Page 20: ...paring for use R S ENV216 20 User Manual 5201 6693 14 05 4 7 Connecting the test receiver The test receiver is connected to the N connector RF OUTPUT 1 by a 50 coaxial cable Connecting the test receiv...
Page 21: ...work using a 1 meter long unshielded wire For devices under test equipped with a permanently attached cable that is longer than 1 meter the cable is laid out in a folded configuration 1 Metal wall at...
Page 22: ...equipment under test If the housing is made from non conductive material metal foils are wound round the handles and a 60 mm wide foil is wound round the point on the housing that can also be gripped...
Page 23: ...nnection cable running separately to the hand simulation 5 Two Line V Network 6 Metallic collar capable being gripped 7 Test receiver A Power supply connection B Reference ground connection low induct...
Page 24: ...manually prior to the remote operation is restored The settings are not saved if the device is disconnected from mains supply instead the default status L1 ON and 150 kHz HP OFF is always restored whe...
Page 25: ...Remote control R S ENV216 25 User Manual 5201 6693 14 05 12 GND 13 TTL 5V 14 N 15 L1 18 FLT...
Page 26: ...S ENV216 from the mains supply 2 Let the R S ENV216 cool down to ambient temperature 3 Check the following parts for damage a Housing displays carrying handles air vents b Device connection cables co...
Page 27: ...resistance The resistance of the PE connection must be checked by a measurement according to Figure 6 1 The test current is set to 10 ADC Figure 6 1 Measuring the resistance of the PE connection with...
Page 28: ...d when you need it For quick expert help with any Rohde Schwarz product contact our customer sup port center A team of highly qualified engineers provides support and works with you to find a solution...
Page 29: ...in a way that causes minimum environmental impact Disposing electrical and electronic equipment A product that is labeled as follows cannot be disposed of in normal household waste after it has come t...
Page 30: ...cific Connector Models Germany as well as Australia Finland the Nether lands Norway Russia Sweden Occasionally Portugal Spain R S ENV216 3560 6550 12 UK as well as Ireland Hong Kong Malaysia Singa por...
Page 31: ...Connector Models R S ENV216 31 User Manual 5201 6693 14 05 China as well as Australia and New Zealand R S ENV216 3560 6550 15 USA as well as Canada Japan Korea Taiwan Mex ico Central America R S ENV2...