Measurement basics
R&S
®
NRX
49
User Manual 1178.5566.02 ─ 09
a) Press the [Preset] key.
b) Tap "Preset".
Chapter 8, "Saving and recalling settings"
2. Depending on the power sensor and the measurement conditions, consider to zero
the power sensor:
Execute zeroing:
Note:
Turn off all measurement signals before zeroing. An active measurement
signal during zeroing causes an error.
a) Switch off the power of the signal source.
b) Press the [Zero] key.
c) Tap "Zero All Sensors".
3. Configure the measurement.
a) Open the "Measurement Settings" dialog, as described in
b) Select the "Measurement Type", for example "Continuous Average".
c) Tap "Quick Setup"
> "Auto Set".
4. Switch on the signal source.
The measurement starts, and the result is displayed in dBm.
5. If necessary, perform further settings.
5.4
Limit violation
If a measured value violates the set limits, it is highlighted in red.
► To change the limit settings, tap the displayed limit values. See also (2) in
.
Limit violation