Remote control commands
R&S
®
NRX
402
User Manual 1178.5566.02 ─ 09
SYSTem:FWUPdate:STATus?
Queries the result of the firmware update, returned as a readable string.
Further information.
●
●
Example:
SYST:FWUP:STAT?
Query
"Success"
Response
Usage:
Query only
13.14.3.2
Updating a connected power sensor
Prerequisite: The power sensor is connected as described in
............................................................................. 402
SENSe<Sensor>:SYSTem:FWUPdate:STATus?
................................................................402
SENSe<Sensor>:SYSTem:FWUPdate
<fwudata>>
Loads new operating firmware into the selected power sensor. Rohde
&
Schwarz pro-
vides the update file. For further details, see
If you want to integrate a firmware update function in an application, see the example
given in
Chapter 13.14.3, "Firmware update"
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Setting parameters:
<<fwudata>>
<block_data>
See
Usage:
Setting only
SENSe<Sensor>:SYSTem:FWUPdate:STATus?
Queries the result of the firmware update, returned as a readable string.
Further information.
●
●
System information and configuration