Remote control commands
R&S
®
NRX
390
User Manual 1178.5566.02 ─ 09
................................................................................................391
.................................................................................. 392
CALibration<Sensor>:TEST?
Requires a thermal power sensor, equipped with an additional test heater.
Queries the power difference in W when measuring with the external heater enabled
and disabled.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Query only
CALibration<Sensor>:TEST:DEViation?
Requires a thermal power sensor, equipped with an additional test heater.
Queries the relative deviation of the current power difference from the reference value
stored in the calibration data set. The relative deviation is calculated as follows:
<relative
deviation>
= (<current
power
difference>
/ <reference
value>) - 1
with
<current power difference>:
CALibration<Sensor>:TEST:REFerence?
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Query only
CALibration<Sensor>:TEST:REFerence?
Requires a thermal power sensor, equipped with an additional test heater.
Queries the reference value of the power difference in W when measuring with the
external heater enabled and disabled. The reference value is determined during the
calibration process and stored in the calibration data set.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Query only
Running selftests