![Rohde & Schwarz 1424.7805K02 User Manual Download Page 379](http://html1.mh-extra.com/html/rohde-and-schwarz/1424-7805k02/1424-7805k02_user-manual_1477791379.webp)
Remote control commands
R&S
®
NRX
379
User Manual 1178.5566.02 ─ 09
Sets the termination resistance of the external trigger signal that is supplied at the
Trig
In
/
Out
2 connector. Choose a setting that fits the impedance of the trigger source
to minimize reflections on the trigger signals.
Suffix:
<undef>
.
1 to n
No suffix required.
Parameters:
<impedance>
HIGH | LOW
*RST:
HIGH
Manual operation:
See
OUTPut:LIMit:FAIL
<mode>
is set to:
LIM
,
FLIMit
or
RLIMit
Sets the fail voltage that is output if a measured value causes a limit violation.
Parameters:
<mode>
LOW | HIGH
HIGH
Output voltage of 5
V.
LOW
Output voltage of 0
V.
*RST:
LOW
Manual operation:
See
OUTPut:LIMit:FEED:INDex
<index>
LIM
or
FLIM
or
RLIM
is set.
Sets the measurement that is monitored.
Parameters:
<index>
Manual operation:
See
"Measurement for Limit Output"
OUTPut:MODE<output>
<mode>
Sets the functionality of the Out
1
/
Trig
Out and Trig
In
/
Out
2 connectors.
Configuring the analog signal ouput and the trigger input/output