
Remote control commands
R&S
®
NRX
367
User Manual 1178.5566.02 ─ 09
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:BURSt:MODE
<mode>
Defines how the average burst power is determined.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<mode>
AUTO | USER
AUTO
Not supported by all power sensors.
The power sensor automatically recognizes the duty cycle of the
burst series and calculates the average burst power from this
duty cycle and the average power. Set an appropriate video
bandwidth using
CHANnel<Channel>]:NRT:VBWidth[:VALue]
USER
Define the duty cycle by:
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:
BURSt:PERiod
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:
BURSt:WIDTh
The R&S NRX calculates the average burst power from these
values.
*RST:
AUTO
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:BURSt:PERiod
<value>
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:BURSt:MODE
USER
is set.
Sets the burst period.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor or 1 = forward, 2 =
reflection (reverse)
Parameters:
<value>
Range:
0.0 to 1.0
*RST:
0.1
Default unit: s
Configuring sensors