
Sensor configuration
R&S
®
NRX
131
User Manual 1178.5566.02 ─ 09
...............................................................................................131
Offset Reference Plane
Selects the power sensor port to which the measurement results are referred to.
"Source"
Source connector of the R&S
NRT
‑
Zxx power sensor
"Load"
Load connector of the R&S
NRT
‑
Zxx power sensor
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:OFFSet:
RPLane
INPut<Sensor>:PORT:POSition
Offset
Considers the transmission loss in a cable that connects the desired measurement
point, set by
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:OFFSet:
STATe
CALCulate<Measurement>[:CHANnel<Channel>]:CORRection:OFFSet[:
MAGNitude]
INPut<Sensor>:PORT:OFFSet
Modulation
Sets a communication standard for the modulation correction to reduce systematic
deviations occurring in power measurements.
"Off"
Disabled.
NRT measurement type