![Rohde & Schwarz 1410.4409.02 User Manual Download Page 57](http://html1.mh-extra.com/html/rohde-and-schwarz/1410-4409-02/1410-4409-02_user-manual_1477786057.webp)
R&S
RT
‑
ZA15 External Attenuator
R&S
®
RT
‑
ZD10/20/30
34
User Manual 1410.4550.02 ─ 01
3.5
Maximum Input Voltage
The maximum input voltage of the R&S
RT
‑
ZA15 external attenuator is specified
as the maximum voltage between each signal socket and ground.
Shock hazard caused by high voltages
To avoid electric shock and personal injury, and to prevent damage to the
probe or any other products connected to it, observe the following instruc-
tions:
●
The maximum input voltage for DC signals is ±70 V between each sig-
nal socket and the ground socket.
●
The maximum effective input voltage for AC signals is 33 V (eff)
between each signal socket and the ground socket. The maximum tran-
sient peak voltage is ±46 V.
The maximum input voltage is derated for higher frequencies. Refer to the
R&S
RT
‑
Z data sheet for further information.
Maximum Input Voltage