Operating the R&S
QAR
R&S
®
QAR
41
User Manual 1178.7200.02 ─ 09
Table 5-1: Keyboard shortcuts
Ctrl-S
Saves the currently displayed measurement results.
"Managing measurement results"
for details.
Ctrl-O
Opens a dialog box to load measurement results.
"Managing measurement results"
for details.
Ctrl-Q
Closes the application.
F3
Initiates a reflection normalization
F4
Initiates a transmission normalization
F5
Initiates a measurement.
5.1.3
Measurements
The R&S
QAR provides several measurement types: reflection measurements and
transmission measurements. Transmission measurements require the optional trans-
mitter module. Both measurements usually consist of two steps, normalization and the
actual measurement.
Measuring a DUT
1. Set up the normalization.
2. Start the normalization with "Normalize Reflection" (or the F3 key) and "Normalize
Transmission" (or the F4 key).
For details on how to normalize the measurement, see
3. Place the DUT in front of the R&S
QAR, for example on the optional mounting
table.
4. Start the reflection and transmission measurement with "Measure" (or the F5 key).
The measurement itself lasts less than a second. The results are displayed after a
few seconds.
For details on how to run the measurement, see
Chapter 5.1.3.2, "Reflection Mea-
Chapter 5.1.3.3, "Transmission Measurement"
......................................................................................... 43
....................................................................................43
5.1.3.1
Normalization
Normalization is a separate measurement that determines systematic, reproducible
measurement deviations. Determining these measurement deviations allows you to
Radome Measurements