
Remote control commands
R&S
®
RTM3000
494
User Manual 1335.9090.02 ─ 09
17.6.4.6
Runt search configuration
....................................................................................494
SEARch:TRIGger:LEVel:RUNT:LOWer
.............................................................................494
SEARch:TRIGger:LEVel:RUNT:UPPer
............................................................................. 494
..................................................................................... 494
...................................................................................... 495
.......................................................................................495
SEARch:TRIGger:RUNT:POLarity
<Polarity>
Indicates the polarity of a the runt to be searched for.
Parameters:
<Polarity>
POSitive | NEGative | EITHer
*RST:
POS
SEARch:TRIGger:LEVel:RUNT:LOWer
<LowerLevel>
Sets the lower voltage threshold for runt detection. A positive runt crosses the lower
level twice without crossing the upper level.
Parameters:
<LowerLevel>
Range:
Depends on vertical scale and LSB value.
*RST:
400 mV
Default unit: V
SEARch:TRIGger:LEVel:RUNT:UPPer
<UpperLevel>
Sets the upper voltage threshold for runt detection. A negative runt crosses the upper
level twice without crossing the lower level.
Parameters:
<UpperLevel>
Range:
Depends on vertical scale and LSB value.
*RST:
600 mV
Default unit: V
SEARch:TRIGger:RUNT:RANGe
<Range>
Sets how the measured pulse width is compared with the given limit(s).
.
.
Parameters:
<Range>
LONGer | SHORter | WITHin | OUTSide
LONGer
Finds pulses longer than the given width.
SHORter
Finds pulses shorter than the given width.
Waveform analysis