Command Reference
R&S
®
ZNA
1242
User Manual 1178.6462.02 ─ 12
Manual operation:
See
SOURce<Ch>:POWer<Pt>:ALC:TPOWer
<Value>
Defines the target value the ALC at port <Pt> in channel <Ch> stabilizes.
The related receiver whose power .
Suffix:
<Ch>
.
<Pt>
Port number
Parameters:
<Alc target power for
b or b' receiver>
Default unit: dBm
Manual operation:
See
"ALC: Meas Port/ALC: Meas Receiver/ALC: Target Power"
SOURce:POWer:CONVerter
Allows you to specify source power information for connected mm-wave converters
(with R&S
ZNA-K8).
SOURce:POWer<PhyPt>:CONVerter:TRANsfer:DESCription
<Power control mode>
For the converter connected to physical VNA port <PhyPt>, sets/queries whether a lev-
eling dataset shall be used.
Suffix:
<PhyPt>
.
Physical port number
Parameters:
<Power control
mode>
NONE | DSET
NONE
Don't use a leveling dataset.
DSET
Use a leveling dataset.
This can be either the factory leveling dataset a leveling dataset
created with the converter leveling tool (see
POWer<PhyPt>:CONVerter:TRANsfer:DEST
Options:
R&S
ZNA-K8
Manual operation:
See
SOURce:POWer<PhyPt>:CONVerter:TRANsfer:DEST
<DataSetMode>
For the frequency converter connected to physical port <PhyPt>, sets/queries whether
the factory leveling dataset or a user-defined leveling dataset shall be used.
Only takes effect if
SOURce:POWer<PhyPt>:CONVerter:TRANsfer:DESCription
is set to
DSET
.
SCPI Command Reference