[SENSe<Ch>:]SEGMent<Seg>:PORT<PortId>:POWer[:LEVel]
<IntSourcePortPow>
Defines a port-specific power level for port <PortId> in sweep segment <Seg> of chan-
nel <Ch>.
Requires that a port-specific power level has been previously enabled using
[SENSe<Ch>:]SEGMent<Seg>:PORT<PortId>[:STATe]
.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
<PortId>
Port number
Parameters:
<IntSourcePortPow> Power level for port <PortId> in sweep segment <Seg> of chan-
nel <Ch>
Default unit: dBm
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:PORT<PortId>[:STATe]
<Boolean>
Enables a port-specific power level (base power) for port <PortId> in segment <Seg>
of channel <Ch>.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
<PortId>
Port number
Parameters:
<Boolean>
OFF (0)
No port-specific power level for port <PortId> in segment <Seg>
of channel <Ch>.
If
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]:CONTrol
is set to ON, the segment power level
LEVel][:IMMediate][:AMPLitude]
ON (1)
The port-specific power level
segment <Seg> of channel <Ch>.
*RST:
OFF
Manual operation:
See