Command Reference
R&S
®
ZNA
1124
User Manual 1178.6462.02 ─ 12
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer. This numeric suffix is ignored; the
active port is determined by the active trace.
Setting parameters:
<Activate>
ONCE
Applies the "Auto Length and Loss" function.
Example:
*RST; :CORR:LOSS:AUTO ONCE
Reset the instrument and apply the "Auto Length and Loss"
function to the default trace (Trc1 in channel 1).
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:FREQuency
<RefFreq>
Defines the reference frequency for the frequency-dependent part of the offset loss
(
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:OFFSet
).
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer
Parameters:
<RefFreq>
Reference frequency
Range:
Frequency range of the analyzer model.
Increment: 1 MHz
*RST:
1000000000 Hz (= 1 GHz)
Default unit: Hz
Example:
See
[SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth
Manual operation:
See
"Loss at DC / Loss at Freq / Freq for Loss"
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:OFFSet
<OffsetLoss>
Defines the frequency-independent part (DC value) of the offset loss.
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer
Parameters:
<OffsetLoss>
Frequency-independent part of the offset loss
Range:
-200 dB to +200 dB
Increment: 0.001 dB
*RST:
0 dB
Default unit: dB
SCPI Command Reference