10.5.2.1
CALCulate Commands
The
CALCulate...
commands perform post-acquisition data processing. Functions
in the
SENSe
subsystem are related to data acquisition, while the
CALCulate
subsys-
tem operates on the data acquired by a
SENSe
function.
CALCulate:CLIMits...
The
CALCulate:CLIMits...
commands control the composite limit check.
CALCulate:CLIMits:FAIL?
Returns a 0 or 1 to indicate whether or not a global, composite limit check on several
traces has failed.
Since V2.20 of the R&S
ZNL/ZNLE FW the result is automatically recalculated when-
ever a relevant setting is changed, i.e. a subsequent query will return the updated limit
violation state.
Example:
*RST; CALC:LIM:CONT 1 GHZ, 2 GHZ
Define an upper limit line segment in the stimulus range
between 1 GHz and 2 GHz, using default response values.
CALC:LIM:STAT ON; FAIL?
Switch the limit check on and query the result.
CALC:CLIM:FAIL?
Query the result for the composite limit check. As only one trace
is tested, the response should be equal to the previous
response.
Usage:
Query only
Manual operation:
See
CALCulate:DATA...
The
CALCulate:DATA...
commands provide access to the results of a measure-
ment.
Data format
The trace data is transferred in either ASCII or block data (REAL) format, depending
on the
setting. If block data format is used, it is recommended to
select EOI as a receive terminator (
SYSTem:COMMunicate:GPIB[:SELF]:
EOI
).
VNA Command Reference