Command Reference
R&S
®
ZNL/ZNLE
896
User Manual 1178.5966.02 ─ 07
STATus:QUEStionable:PTRansition
<BitPattern>
STATus:QUEStionable:INTegrity:PTRansition
<BitPattern>
STATus:QUEStionable:INTegrity:HARDware:PTRansition
<BitPattern>
STATus:QUEStionable:LIMit<Lev>:PTRansition
<BitPattern>
Configures the positive transition filters of the
QUEStionable...
status registers. If a
bit is set, a 0 to 1 transition in the corresponding bit of the associated condition register
causes a 1 to be written in the associated bit of the corresponding event register.
See also
Chapter 10.4.4.5, "Reset Values of the Status Reporting System"
Suffix:
<Lev>
.
Selects one of the two
QUEStionable:LIMit
registers; see
"STATus:QUEStionable:LIMit<1|2>"
Parameters:
<BitPattern>
Range:
0 to 65535 (decimal representation)
*RST:
n/a
Example:
STAT:QUES:LIM2:PTR 6
Set bits no. 1 and 2 of the
QUEStionable:LIMit2:PTRansition
register
STATus:QUEue[:NEXT]?
Queries and at the same time deletes the oldest entry in the error queue. Operation is
identical to that of
.
The entry consists of an error number and a short description of the error. Positive
error numbers are instrument-dependent. Negative error numbers are reserved by the
SCPI standard; see section
Chapter 11.1, "Error Messages and Troubleshooting"
Example:
STAT:QUE?
Query the oldest entry in the error queue.
0,"No error"
is
returned if the error queue is empty.
Usage:
Query only
10.5.2.16
SYSTem Commands
The
SYSTem...
commands provide functions that are not related to instrument per-
formance, such as functions for general housekeeping and functions related to global
configurations.
SYSTem:COMMunicate...
The
SYSTem:COMMunicate...
commands provide remote control settings and con-
figure remote (external) devices controlled by the R&S
ZNL/ZNLE.