Remote Commands
R&S
®
Spectrum Rider FPH
439
User Manual 1321.1011.02 ─ 11
[SENSe:]SWEep:TIME
<SweepTime>
This command defines the sweep time.
If you set a sweep time with this command, the R&S Spectrum Rider decouples the
sweep time from the span and the resolution and video bandwidths.
Parameters:
<SweepTime>
Range:
Specified in the datasheet.
*RST:
- (AUTO is set to ON)
Default unit: s
Example:
SWE:TIME 10s
Sets the sweep time to 10 s.
Triggering Measurements
The following commands set up trigger conditions if you are using a trigger for the
measurement.
List of Commands
.......................................................................................... 439
........................................................................................439
TRIGger[:SEQuence]:SLOPe
<TriggerSlope>
This command selects the slope of the trigger signal.
The trigger slope applies to all trigger sources.
Parameters:
<TriggerSlope>
POSitive | NEGative
*RST:
POSitive
Example:
TRIG:SLOP NEG
TRIGger[:SEQuence]:SOURce
<TriggerSource>
This command selects the trigger source.
Parameters:
<TriggerSource>
IMMediate | EXTernal | VIDeo | IQPower | GATed
IMMediate
Selects Free Run measurements.
EXTernal
Selects an external trigger.
VIDeo
Selects the video trigger.
IQPower
Selects an IQ Power trigger.
Remote Control - Commands
www.allice.de