Remote Commands
R&S
®
Spectrum Rider FPH
394
User Manual 1321.1011.02 ─ 11
[SENSe:]FILTer:DEMPhasis[:STATe]
<State>
This command switches the FM deemphasis filter on or off.
Parameters:
<State>
ON | OFF
ON
Turns on the FM deemphasis filter.
OFF
Turns off the FM deemphasis filter.
*RST:
OFF
Example:
FILT:DEMP OFF
Switches off the FM deemphasis filter.
13.6.3.5
Defining the Sweep Mode
The following chapter describes command that define the sweep mode and query
measurement time.
List of commands
............................................................................................. 394
[SENSe:]ADEMod:MTIMe?
This command queries the measurement time in the analog modulation.
Example:
ADEM:MTIM?
Queries the measurement time in the analog modulation
Usage:
Query only
INITiate[:IMMediate]
This command initiates a new measurement sequence.
With sweep count > 0 or average count > 0, this means a restart of the indicated num-
ber of measurements. With trace functions MAXHold, MINHold and AVERage, the pre-
vious results are reset on restarting the measurement.
In single sweep mode, synchronization to the end of the indicated number of measure-
ments can be achieved with the command *OPC, *OPC? or *WAI. In continuous–
sweep mode, synchronization to the sweep end is not possible since the overall mea-
surement never ends.
This command is an event and therefore has no query and no *RST value.
Remote Control - Commands