Remote Commands
R&S
®
Spectrum Rider FPH
389
User Manual 1321.1011.02 ─ 11
[SENSe:]FREQuency:CENTer:STEP
<Stepsize>
This command defines the center frequency step size.
Parameters:
<Stepsize>
Range:
1 Hz to fmax
*RST:
- (AUTO 0.1*SPAN is switched on)
Default unit: MHz
Example:
FREQ:CENT:STEP 120MHz
Defines a CF step size of 120 MHz.
13.6.3.2
Defining the Vertical Axis
The following chapter describes command necessary to define the vertical scale such
as the x-axis deviation per division and settings affecting the amplitude of the measure-
ments.
List of commands
CALCulate<1...2>:MARKer<1...6>:FUNCtion:LEVel:ONCE
.................................................389
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:PDIVision
.......................................... 390
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:RLEVel
..............................................390
............................................................................................... 390
............................................................................................... 391
...............................................................................................391
CALCulate<1...2>:MARKer<1...6>:FUNCtion:LEVel:ONCE
This command adjusts the reference level to the measured signal power.
This automatic routine makes sure that the signal power level does not overload the
R&S Spectrum Rider or limits the dynamic range by too small an S/N ratio.
To determine the best reference level, the R&S Spectrum Rider aborts current mea-
surements and performs a series of test sweeps. After it has finished the test, it contin-
ues with the actual measurement.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
CALC:MARK:FUNC:LEV:ONCE
Initiates an automatic level adjust routine.
Usage:
Event
Remote Control - Commands