Remote Commands
R&S
®
Spectrum Rider FPH
330
User Manual 1321.1011.02 ─ 11
Example:
SWE:POIN?
Returns the number of sweep points.
[SENSe:]SWEep:TIME
<SweepTime>
This command defines the sweep time.
If you set a sweep time with this command, the R&S Spectrum Rider decouples the
sweep time from the span and the resolution and video bandwidths.
Parameters:
<SweepTime>
Range:
Specified in the datasheet.
*RST:
- (AUTO is set to ON)
Default unit: s
Example:
SWE:TIME 10s
Sets the sweep time to 10 s.
[SENSe:]SWEep:TIME:AUTO
<State>
This command couples and decouples the sweep time to the span and the resolution
and video bandwidths.
Parameters:
<State>
ON | OFF
*RST:
ON
Example:
SWE:TIME:AUTO ON
Switches on the coupling to frequency span and bandwidths.
Triggering Measurements
The following commands set up trigger conditions if you are using a trigger for the
measurement.
List of commands
....................................................................................331
.....................................................................................331
........................................................................................ 331
TRIGger[:SEQuence]:LEVel:VIDeo
.................................................................................. 331
.......................................................................................... 332
........................................................................................332
[SENSe:]SWEep:EGATe
<GateTrigger>
This command turns a gated trigger setting mode on and off.
Parameters:
<GateTrigger>
ON | OFF
*RST:
OFF
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH