Spectrum Analyzer Mode
R&S
®
Spectrum Rider FPH
204
User Manual 1321.1011.02 ─ 11
1. Press [MARKER] key.
2. Select "Marker Function" softkey.
3. Select the "n dB Down" menu item.
The R&S Spectrum Rider displays two temporary markers on the left and on the
right of the reference marker, "M1". It also shows the bandwidth between the n dB
down markers. You can then adjust the distance of the temporary markers.
4. Select "Marker Function" softkey.
5. Select the "n dB Down" menu item.
The R&S Spectrum Rider opens an input field.
6. Enter a different distance of 6 dB.
The R&S Spectrum Rider again shows the temporary markers, this time with a
broader bandwidth.
Demodulating signals
Demodulating signals
is activated, the R&S Spectrum Rider disables the demodulator function
in the
.
The R&S Spectrum Rider features an AM and FM demodulator to demodulate and
monitor audio signals. It demodulates the signal at the marker frequency.
You can listen to the demodulated signal with the internal speaker or headphones that
you can connect to the 3.5 mm headphone jack on the top of the R&S Spectrum Rider.
See
Chapter 3.2.2.3, "Headphone Jack"
Configuring Spectrum Measurements
www.allice.de