Remote control commands
R&S
®
NRPxxTWG
47
User Manual 1178.8371.02 ─ 08
Sets the delay between the artificial trigger event and the beginning of the actual mea-
surement
Parameters:
<delay>
Range:
0.1 to 5.0
*RST:
2.0
Default unit: Seconds
TRIGger:ATRigger:EXECuted?
Queries the number of measurements that were triggered automatically since
was set to
ON
.
In normal scalar measurements, this number can only be 0 or 1. If a buffered measure-
ment was executed, this number indicates how many results in the returned array of
measurement data were executed without a real trigger event.
Usage:
Query only
TRIGger:ATRigger[:STATe]
<state>
Controls the automatic trigger function. If enabled, an artificial trigger is generated if the
delay time has elapsed after the measurement start and no trigger event has occurred.
The delay time is set using
Parameters:
<state>
*RST:
OFF
TRIGger:COUNt
<count>
Sets the number of measurement cycles to be performed when the measurement is
started using
.
This number equals the number of results that can be obtained from the sensor after a
single measurement. As long as the defined number of measurements is not executed,
the sensor automatically initiates another measurement internally when the current
result is available.
This command is particularly useful in conjunction with buffered measurements. For
example, to fill a buffer with a predefined size with measurements that have been trig-
gered externally or by
without having to start the measurement multiple times.
Parameters:
<count>
Range:
1 to 8192
*RST:
1
Example:
See
Chapter 8.2, "Performing a buffered continuous average
Controlling the measurement