Module Replacement
Radio Unit R&S
CRTU-RU
1140.0050.82 3.36
E-4
Replacing the TR-CORRECTION MODULE in the RXTX BOARD1/2
(see chapter 5,
spare part list, Current No. 360/380
and
explosion drawing 1100.1733.01 DF sheet 3
)
The TR-CORRECTION MODULE is incorporated in the RXTX BOARD1/2 as sandwich module. For
replacement proceed as follows:
Opening the instrument and removing the TR-CORRECTION MODULE
¾
Switch off the instrument, pull the mains plug and unscrew the rear-panel feet.
¾
Place the instrument onto the front carrying handles, push the tube upwards and take off.
¾
After unscrewing the two countersunk screws at the top left instrument frame, slightly lift the cover at
the instrument top, slightly push to the right and lift off.
¾
Place the instrument to the left side and unscrew the two RF SMA coax cables at the bottom of the
RXTX BOARD1/2 using a special wrench. Likewise, pull off the MMCX plug-in cable at the bottom.
¾
Pull off all the MMCX plug-in cables (using a pull-out tool
∗
) from the top of the RXTX BOARD1/2 and
pull off the module towards the top using the two levers on the right and left.
¾
Unscrew one screw of each cooling plate, pull off the cooling plates towards the side and then
unscrew all screws of the RXTX aluminum cover shells. Then the cover shells of the top and bottom
can be removed.
¾
Pull off the sandwich module TR-CORRECTION MODULE (using a pull-out tool
∗
) from the RXTX
BOARD1/2.
Installing the new TR-CORRECTION BOARD and completing the instrument
¾
Insert the new TR-CORRECTION MODULE.
¾
Reinsert the cover shells in the reverse order and fasten with screws. Attach the cooling plates again.
¾
Insert the RXTX BOARD1/2 into the instrument in the reverse order and carefully connect all the
MMCX plug-in cables and RF-SMA coax cables.
¾
Complete the instrument without causing damage to the cables.
¾
Connect the instrument to the mains.
¾
Connect the external keyboard to an USB connector of the CRTU Radio Unit.
¾
Switch on the CRTU Radio Unit.
Automatic module data adjustment
¾
Boot up within BootMagic
menu into “DOS OPSW / Service” or “Service / WCDMA”.
¾
During startup observe the display.
¾
When three BEEPs can be heard, press the MENU SELECT key.
¾
The version manager is started (see also chapter 4, Firmware update). The display includes the
menu item FIRMWARE UPDATE AFTER BOARD CHANGE.
¾
Press the softkey to the left of the above mentioned menu item.
The automatic module data adjustment is performed under software control, firmware updates for
microprocessors and programmable devices being performed as well. This may take a few minutes.
∗
available in CMU-Service Kit CRTU-Z3 (1139.0509.02)
Summary of Contents for 1138.4000
Page 4: ......
Page 6: ......
Page 16: ......
Page 18: ......
Page 64: ......
Page 66: ......
Page 136: ......
Page 148: ......
Page 151: ...1140 0050 82 5 3 E 4 Overview of Spare Parts for CRTU Radio Unit...
Page 152: ......
Page 161: ...1140 0050 82 5 13 E 3 Drawings of all Radio Unit R S CRTU RU Modules...
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ...1140 0050 82 6 1 E 3 Circuit Diagrams for Radio Unit R S CRTU RU...
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......
Page 178: ......
Page 179: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 183: ......
Page 184: ......
Page 185: ......
Page 186: ......
Page 187: ......
Page 188: ......
Page 189: ......
Page 190: ......
Page 191: ......
Page 192: ......
Page 193: ......
Page 194: ......
Page 195: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......