Ground Sense Wire
Power Cord
metal part of handle
Device Under
Test (DUT)
DUT with 3 Prong Plug
(If you are testing a DUT with other than
a three prong grounded power cord, see
diagram below titled : “SET-UP TESTING
2-WIRE HIPOT”)
WHAT IS NEEDED FOR FOR SELF TEST AND BASIC DEVICE TESTING:
(The ground sense wire has a banana plug
which inserts into the “Chassis Ground
Sense post on the front panel of the
ROD-L unit and allegator clip to be
connected to the DUT)
(The power cord connects to the rear
panel of the ROD-L unit and to any
grounded three-prong wall socket. Be
sure to set the line voltage on rear panel
to correct voltage i.e. 110 or 220)
1.a
Make sure the line voltage
switch on rear panel is set to
the correct value for the wall
outlet (i.e. 110 or 220).
1.b
Connect the POWER
CORD to the rear of the
ROD-L Unit and plug into
any three-prong grounded
wall outlet.
1.c
Turn on the Line Power on
the front panel of the ROD-
L Unit.
1.d
Insert banana end of
GROUND SENSE WIRE into
the Chassis Ground Sense
post on the front panel of
ROD-L unit.
1.e
Connect allegator clip of
ground sense wire to the
metal portion of the ROD-L
unit handle. Green Ready
light will light.
1.f
Press “Start.”
After you press start, the “Testing” light will light for a brief period of time and then shut off.
This completes the self test.
Disconnecting the Ground Sense Wire from the Chassis Ground Sense post or the metal part of handle will cause
the test to fail. The red fail light will light, the audible alarm will sound and the unit will shut down. If this happens,
press the Reset button to turn off the alarm.
Make sure your test parameters are set correctly on the rear panel. If you are unsure or need to set any test
parameters, see instructions on the next page or refer to the manual.
2.a
Plug the three-prong plug of
the DUT into the front panel
receptacle of the ROD-L
tester
2. BASIC DEVICE TEST
2.b
Connect the allegator clip of
the Ground Sense Wire to
the DUT. Green Ready light
should light.
2.c
Press “Start.” Testing light will
light for a brief period of time
and then return to ready
mode if test passes.
2.d
If failure light and audible
alarm turn on, disconnect the
DUT and press the Reset
button on tester to cancel
alarm.
1. SELF TEST