Rocker 167300-11 Instruction Manual Download Page 10

Operation

Operating procedure

1.   Press on/off         switch to activate the pump after making sure all tubing are well connected. 

2.   To setup vacuum by adjusting the regulator        or       .

*  Warning: When pump is used for aqueous solution filtration, please ensure the liquid level of flask 

not exceed safety level.  Failure to comply could result in serious damage to the pump and void the 

warranty. If it occurs, please contact the distributor immediately for service.

1

3

7

Maintenance

How to reset the Prote

TM

 protection device

1.   Remove the moisture trap       and filter cartridge       by rotating clockwise.

2.   Disassemble the filter cartridge       to ensure the buoy is not closed.

3.   Reinstall all components, filter cartridge and moisture trap back.

1. After finishing the experiment, please remove the liquid in the moisture trap and pumping the 

air for at least 2 minutes to withdraw the residual steam.

6

6

5

5

How to replace cartridge

1.   Please replace the cartridge when it becomes dirty or color-changing.

2.   Please remove the moisture trap        by rotating it clockwise.

3.   Replace the cartridge by the same way.

Summary of Contents for 167300-11

Page 1: ...Please read this instruction manual before using this product Rocker 300 300DC 400 410 Rocker 800 801 810 811 Oil Free Vacuum Pump Instruction Manual...

Page 2: ...should not be moved when running If the buoy closes during pump operation it must be manually reset by uninstalling and reinstalling the filter cartridge 6 The filter cartridge is used to absorb mois...

Page 3: ...Rocker 300 Main Part Diagram 1 2 3 4 5 6 7 Position 1 2 3 4 5 6 7 Designation On Off Switch Muffler Air Outlet Air Inlet Vacuum Gauge ProteTM Filter Cartridge Moisture Trap Drain Valve Regulator...

Page 4: ...300DC Main Part Diagram 1 Position 1 2 3 4 5 6 7 8 Designation On Off Switch Muffler Air Outlet Air Inlet Vacuum Gauge ProteTM Filter Cartridge Moisture Trap Drain Valve Regulator Power Cord 8 3 4 5...

Page 5: ...400 410 Main Part Diagram 1 2 3 4 5 6 7 Position 1 2 3 4 5 6 7 Designation On Off Switch Muffler Air Outlet Air Inlet Vacuum Gauge ProteTM Filter Cartridge Moisture Trap Drain Valve Regulator 1 2 3 4...

Page 6: ...Rocker 800 Main Part Diagram Position 1 2 3 4 5 Designation Muffler Air Outlet Vacuum Gauge Vacuum Regulator Air Inlet On Off Switch 5 4 3 2 1...

Page 7: ...Rocker 801 Main Part Diagram 1 2 3 4 5 6 7 Position 1 2 3 4 5 6 7 Designation On Off Switch Muffler Air Outlet Air Inlet Vacuum Gauge Filter Cartridge Moisture Trap Vacuum Regulator...

Page 8: ...Rocker 810 Main Part Diagram 5 4 3 2 1 Position 1 2 3 4 5 Designation Muffler Air Outlet Vacuum Gauge Vacuum Regulator Air Inlet On Off Switch...

Page 9: ...Rocker 811 Main Part Diagram 1 2 3 4 5 6 7 Position 1 2 3 4 5 6 7 Designation On Off Switch Muffler Air Outlet Air Inlet Vacuum Gauge Filter Cartridge Moisture Trap Vacuum Regulator...

Page 10: ...ice 1 3 7 Maintenance How to reset the ProteTM protection device 1 Remove the moisture trap and filter cartridge by rotating clockwise 2 Disassemble the filter cartridge to ensure the buoy is not clos...

Page 11: ...ker 800 Vacuum pump AC110V 60Hz AC220V 50Hz Rocker 801 Vacuum pump AC110V 60Hz AC220V 50Hz Rocker 810 Vacuum pump AC110V 60Hz AC220V 50Hz Rocker 811 Vacuum pump AC110V 60Hz AC220V 50Hz Moisture trap V...

Page 12: ...Rocker Scientific Co Ltd Tel 886 2 26033311 Fax 886 2 26036622 E mail export rocker com tw https www rocker com tw...