Roche Diagnostics
D-4
Operator’s Manual · Version 1.0
7 General maintenance actions
URISYS 2400
Maintenance menu
Maintenance menu
Using the [Maintenance] screen, the operator can perform general
maintenance tasks, such as washing the liquid handling system, exchanging
the URISYS 2400 cassette, resetting the analyzer, and removing air bubbles.
Maintenance procedures are only possible when the analyzer is in Stand-By
status.
Wash the liquid handling system
Clean the liquid handling system (sample needle, flow cell, and internal
tubing) every day, before the analyzer is switched off. Use a recommended
washing solution, for example, MultiClean, and the “Wash” procedure of the
analyzer to prevent contamination and urinary deposits in the liquid handling
system. Change the washing solution daily.
There are two methods for executing the wash procedure:
o
Use the wash rack on the rack input line.
o
Use the washing solution at the STAT position.
a
To use the wash rack on the input line
1
Register the wash rack ID number to be used for the wash procedure.
2
Place a sample tube containing 10 ml of washing solution at position 1 of
the registered rack, and set the rack on the rack input line.
3
If the analyzer is in Operation status, the rack is automatically transferred
to the cross-feed line and its barcode recognized by the barcode reader; the
wash procedure is then performed.
4
If the analyzer is in Sampling Stop or Stand-By status, press <Start> to feed
the wash rack.
5
When the wash procedure is complete, the analyzer moves to Stand-By
status. Remove the wash rack from the rack output line and dispose of the
washing solution.
After a washing procedure, the instrument must be re-started manually by
pressing <Start> in the [Routine Monitor] screen.
“Register the wash rack” on page B-27
Do not use the registered wash rack for routine or control samples.
If the washing cycle is interrupted, carry out the wash procedure again from the
beginning. Otherwise, the remaining detergent in the flow path may damage the
instrument or cause incorrect data.
When the <Stop> button on the [Routine Monitor] screen is pressed during the
wash procedure, a [Warning] window alerts the operator to the risks of interrupting
the washing process.
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......