Roche Diagnostics
Operator’s Manual · Version 1.0
B-27
URISYS 2400
3 Software
Installation
Password
The operator can change the password used to gain access to the
[Sieve & Abnormal Setting], [Set Range], [Set Reflectance], and [Default]
screens. The password is identical in each case. Only numerical characters can
be used.
a
To change the password
1
Press <Password> on the [System Parameters] screen. The [Password]
window appears.
2
Enter the old (current) 4-digit password.
3
Enter the new 4-digit password.
4
Re-enter the new password.
5
Press <Enter> after each entry.
6
Press <PgUp> to return to the [System Parameters] screen.
Register the wash rack
The operator can define the ID of the rack to be used for washing. If the
washing procedure is performed using a rack on the rack input line, the
analyzer recognizes the specified rack and performs the washing procedure.
a
To register the wash rack
1
Press <Regist. Wash Rack> on the [System Parameters] screen. The
[Regist.Wash Rack] screen is displayed.
2
Enter the consecutive range of the wash rack ID numbers in the fields
provided. The wash rack ID numbers must consist of five digits, and
overlapping ranges (for example, with routine sample racks) are not
allowed.
3
Press <Enter> after each entry.
4
Press <PgUp> to return to the [System Parameters] screen.
If the 5 digit number of the chosen rack is not known, read and print the rack ID via
the Barcode Check function (see “Checks” on page C-41).
Mark the specified rack as a wash rack. Do not use the wash rack for routine or
control samples; otherwise, the samples will not be measured.
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......