7 Trouble shooting
7-10
Instructions for Use, Roche OMNI C, Rev. 4.0, December 2002
7.3
Electrode status
If a parameter is not ready (not calibrated), the symbol will be crossed out with grey and red
lines (see Abb. 2).
Abb. 2
After you press the parameter button, a status report will give you information about why
the parameter could not be calibrated (see Abb. 3).
Abb. 3
Remedy for the example in Fig. 3
•
Start "Calibration for Ready", see chapter 3, "
Calibration
".
•
Exchange electrode, see chapter 6, "
Maintenance
", section "
Exchanging the Electrodes
and the MCon
."
•
Please see the Reference Manual, chapter 5,
"Error Correction"
, for a detailed descrip-
tion of the short texts!
Summary of Contents for OMNI C
Page 6: ......
Page 44: ...1 Introduction 1 36 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 46: ...2 Specifications 2 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 56: ...2 Specifications 2 10 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 58: ...3 Calibration 3 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 62: ...3 Calibration 3 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 64: ...4 Measurement 4 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 76: ...4 Measurement 4 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 78: ...5 Quality control 5 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 118: ...6 Maintenance 6 24 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 120: ...7 Trouble shooting 7 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 132: ...7 Trouble shooting 7 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 134: ...8 Operating modes 8 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 148: ...8 Operating modes 8 14 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 152: ...Index 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...