![Roche OMNI C Instructions For Use Manual Download Page 109](http://html1.mh-extra.com/html/roche/omni-c/omni-c_instructions-for-use-manual_1472343109.webp)
6 Maintenance
Instructions for Use, Roche OMNI C, Rev. 4.0, December 2002
6-15
•
Open the flap to the labelled capillary position (completely open position).
•
Remove the sample drop cup and clean it according to local regulations (see Fig. 17/1).
•
Open the bottle compartment cover.
•
Remove the red tube connector from the wash plate (see Fig. 17/2)
•
Push the wash plate down (to unlock) and pull it out (see Fig. 17/3). The wash plate is
locked when the plug is sticking out and cannot be removed.
Fig. 17
•
It is best to clean the wash plate under running water. Then, dry well.
•
Then push the wash plate back in completely and attach the tube plug.
•
Push the sample drip tray in.
•
Close the flap.
•
Close the bottle compartment cover.
6.6.3 Replacing printer paper
TIP: The printer paper is heat sensitive on one side only. Please make sure that you insert
the paper roll correctly.
•
Open the printer cover.
•
Remove the empty cardboard roll.
•
Cut off, at a right angle, the start of the paper.
•
Place the new paper roll into the holder.
•
Make sure that the printer lever is in the "down" position (see Fig. 18).
•
Feed in the beginning of the paper according to the sticker on the inside of the printer
cover (see Fig. 18).
Fig. 18
(1)
(2)
(3)
printer lever
Summary of Contents for OMNI C
Page 6: ......
Page 44: ...1 Introduction 1 36 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 46: ...2 Specifications 2 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 56: ...2 Specifications 2 10 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 58: ...3 Calibration 3 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 62: ...3 Calibration 3 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 64: ...4 Measurement 4 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 76: ...4 Measurement 4 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 78: ...5 Quality control 5 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 118: ...6 Maintenance 6 24 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 120: ...7 Trouble shooting 7 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 132: ...7 Trouble shooting 7 12 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 134: ...8 Operating modes 8 II Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 148: ...8 Operating modes 8 14 Instructions for Use Roche OMNI C Rev 4 0 December 2002...
Page 152: ...Index 4 Instructions for Use Roche OMNI C Rev 4 0 December 2002...