Roche CoaguChek XS PST Correction Of Service Manual Download Page 2

TP-01142 
page 2 of 5

Issue, continued _________________________________________________________

The current User Manual for your meter does not contain instructions on how to change 
measuring units on the CoaguChek XS PST meter.  This UMDC provides instructions on how 
to ensure your meter is displaying INR results.  Roche is updating the User Manual and the 
package insert labeling for the test strips accordingly.

Actions Required

•  When testing for INR, 

always confirm that “INR” is displayed next to the result after 

each test.  

Reporting your result in a measuring unit other than INR could lead to an 

incorrect assessment of appropriate treatment and potential for harm.

•  Follow the instructions provided in the 

Patient Instructions

 section of this UMDC to 

ensure that the measuring unit result is set to INR.  Refer to these instructions when  
adjusting the date/time in your meter and when the measuring unit needs to be reset to INR.  
Any measurement unit set-up choices other than INR must be authorized by your doctor.

•  If your meter is displaying a result in a different unit than INR and you are unable to reset  

the measuring unit to INR after following the instructions in the 

Patient Instructions

 

section of this UMDC, contact the 

Roche Customer Support Center at 1-800-428-4674

  

for assistance. 

  

•  This Urgent Medical Device Correction (UMDC) serves as supplemental labeling.  File this 

UMDC with your User Manual for future reference.

Patient Instructions ______________________________________________________

Check the Measuring Unit for Your INR Test Result

n

The current result can be changed to a different unit without actually 
changing the settings for measuring units on the meter.  

1.  With the meter powered off, press the 

M

 button and confirm your last test result is 

displayed as“

INR

.”

the measuring unit is displayed next to the result

2.  Press the 

M

 button again to confirm the previous test result also displays as“

INR

.”

Summary of Contents for CoaguChek XS PST

Page 1: ...set to the measuring unit INR There are two additional measuring units on the meter Quick Q and Seconds Sec In the U S the most accepted unit of measure is INR Unless your physician has instructed you to monitor in a unit other than INR results should be reported in INR Roche has been made aware of the following issues The measuring unit of some patient CoaguChek XS PST meters was inadvertently ch...

Page 2: ...ime in your meter and when the measuring unit needs to be reset to INR Any measurement unit set up choices other than INR must be authorized by your doctor If your meter is displaying a result in a different unit than INR and you are unable to reset the measuring unit to INR after following the instructions in the Patient Instructions section of this UMDC contact the Roche Customer Support Center ...

Page 3: ...different order press M until 12 31 00 is reached and press the Set button Now the double digit Year 00 should begin flashing 3 Press the M button until the correct double digit year displays Once the year is correct press the Set button to confirm the year and move to the month The double digit Month 12 should begin flashing 4 Press the M button until the correct double digit month is displayed P...

Page 4: ... or Sec flashing press M until you reach INR and press the Set button to accept This will ensure your test results display in the proper unit of measurement for correct reporting Always ensure INR is displayed when your test result appears unless your doctor has specifically instructed you to measure in a unit other than INR i e Quick or Seconds 2 The Beeper setting should begin flashing which sho...

Page 5: ..._____________________________________________________ Please contact the Roche Customer Support Center Monday through Friday 7 00 a m to 11 00 p m EST at 1 800 428 4674 if you have questions about the information contained in this UMDC This UMDC is being conducted with the knowledge of the U S Food and Drug Administration FDA Adverse events or quality problems experienced with the use of this prod...

Reviews: