Roche CoaguChek XS PST Correction Of Service Manual Download Page 1

TP-01142
12/02/20

CoaguChek

®

 Patients

CoaguChek XS Meters for Patient Self-Testing Use (CoaguChek XS PST) — 
Possible Misinterpretation of Displayed Measuring Units

 

This Urgent Medical Device Correction (UMDC) 

only

 affects patients using:

Product

CoaguChek XS meter

If you are currently using the CoaguChek Vantus meter, please DISREGARD this 
document.

n

The CoaguChek XS PT test strips are not affected by this issue.

Issue ____________________________________________________________________

You are receiving this Urgent Medical Device Correction (UMDC) because records indicate 
that you were shipped a CoaguChek XS PST meter.  This document provides important 
information about your meter.  

Please read this entire document.

n

The CoaguChek XS PST meter is provided to U.S. customers pre-set to the 
measuring unit INR.  There are two additional measuring units on the 
meter: %Quick (%Q) and Seconds (Sec).  In the U.S., the most accepted 
unit of measure is INR.  Unless your physician has instructed you to 
monitor in a unit other than INR, results should be reported in INR.

Roche has been made aware of the following issues:

•  The measuring unit of some patient CoaguChek XS PST meters was inadvertently changed 

by the operator during use from INR to %Quick or Seconds, resulting in a CoaguChek XS 
test result perceived to be different from the laboratory result in INR.  

•  A value displayed in %Quick or Seconds was misinterpreted by a patient as an INR result, 

and the misinterpreted result was significantly higher than the patient’s actual INR.

•  In the above instances, the result produced in %Quick or Seconds was not alleged to  

be inaccurate. 

CoaguChek XS PST

over...

URGENT MEDICAL DEVICE CORRECTION

Roche Diagnostics

Corporation

9115 Hague Road
PO Box 50457
Indianapolis, IN 46250-0457
USA

Summary of Contents for CoaguChek XS PST

Page 1: ...set to the measuring unit INR There are two additional measuring units on the meter Quick Q and Seconds Sec In the U S the most accepted unit of measure is INR Unless your physician has instructed you to monitor in a unit other than INR results should be reported in INR Roche has been made aware of the following issues The measuring unit of some patient CoaguChek XS PST meters was inadvertently ch...

Page 2: ...ime in your meter and when the measuring unit needs to be reset to INR Any measurement unit set up choices other than INR must be authorized by your doctor If your meter is displaying a result in a different unit than INR and you are unable to reset the measuring unit to INR after following the instructions in the Patient Instructions section of this UMDC contact the Roche Customer Support Center ...

Page 3: ...different order press M until 12 31 00 is reached and press the Set button Now the double digit Year 00 should begin flashing 3 Press the M button until the correct double digit year displays Once the year is correct press the Set button to confirm the year and move to the month The double digit Month 12 should begin flashing 4 Press the M button until the correct double digit month is displayed P...

Page 4: ... or Sec flashing press M until you reach INR and press the Set button to accept This will ensure your test results display in the proper unit of measurement for correct reporting Always ensure INR is displayed when your test result appears unless your doctor has specifically instructed you to measure in a unit other than INR i e Quick or Seconds 2 The Beeper setting should begin flashing which sho...

Page 5: ..._____________________________________________________ Please contact the Roche Customer Support Center Monday through Friday 7 00 a m to 11 00 p m EST at 1 800 428 4674 if you have questions about the information contained in this UMDC This UMDC is being conducted with the knowledge of the U S Food and Drug Administration FDA Adverse events or quality problems experienced with the use of this prod...