Meter Setup •
4
69
Operator Lockout
Optional liquid quality controls can be used to ensure
that an operator is performing tests properly on the
CoaguChek XS Pro meter.
The
Operator Lockout
function constrains an operator,
who has been set up in the list, to perform these quality
controls on a regular basis.
1
From the
Main Menu
, touch
Setup
to open the
meter settings.
2
From the
Setup
Menu
, touch
QC Settings
.
3
From the
QC Settings
menu, touch
Operator
Lockout
. If this button is disabled (grayed out),
either no operators have been set up or the
Operator ID
option
has been deactivated.
QC Settings
09:15 am
04/19/2012
QC Lockout
Operator Lockout
STAT Test Con
fi
g.
QC Range
Main Menu
09:15 am
04/19/2012
Control Test
Review Results
Setup
Patient Test
Setup Menu
09:15 am
04/19/2012
ID Setup
QC Settings
Options
Screen
The option of setting up an
Operator Lockout
is availa-
ble only when operator lists are created on the DMS.
For more details see “Data handling”, starting on
page 111. If you are not working with a DMS, only the
QC Lockout
Summary of Contents for CoaguChek XS Pro
Page 1: ...CoaguChek XS Pro P O I N T O F C A R E T E S T I N G Operator s Manual ...
Page 3: ...0 6950540001 02 2015 12 EN CAN Operator s Manual Version 5 0 CoaguChek XS Pro ...
Page 8: ...8 This page intentionally left blank ...
Page 21: ...The CoaguChek XS Pro Meter 2 21 2 The CoaguChek XS Pro Meter A B C D G I J F H K L E ...
Page 110: ...7 Review Results 110 This page intentionally left blank ...
Page 134: ...9 Maintenance and Care 134 This page intentionally left blank ...
Page 138: ...10 Troubleshooting 138 This page intentionally left blank ...
Page 146: ...A Appendix 146 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 ...
Page 150: ...B Appendix 150 This page intentionally left blank ...
Page 154: ...154 This page intentionally left blank ...
Page 155: ......