9
• Maintenance and Care
122
When should the meter be
cleaned and disinfected
■
The meter should be cleaned whenever visibly
soiled.
■
The exterior of the meter and the test strip guide
should be cleaned and disinfected before being
used on another patient.
■
Maintenance cleaning: The use of the disinfecting
cloths may lead to a build-up of residue on the test
strip guide and battery compartment that needs to
be removed. Frequency: whenever residue build-up
is visible or at least once per month.
Follow recommendations from the FDA, CDC and CMC
and your facility’s policies and procedures for infection
control.
1,
2
The FDA recommends that Point of Care testing devices
should be used on one patient only and not shared. If it is
not possible for each individual patient to have a dedi-
cated POC device, the meters must be properly cleaned
and disinfected after every use or before being used on
the next patient, following the guidelines given below.
3
1.
FDA Public Health Notification: Use of Fingerstick Devices
on More than One Person Poses Risk for Transmitting
Bloodborne Pathogens; Initial Communication, (2010).
http://www.fda.gov/MedicalDevicesSafety/
AlertsandNotices/ucm224025.htm
2.
CDC Clinical Reminder: Use of Fingerstick Devices on
More than One Person Poses Risk for Transmitting Blood-
borne Pathogens, (2010).
http://www.cdc.gov/injectionsafety/
FingerstickDevicesBGM.html
3.
Healthcare Infection Control Practices Advisory Committee
(HICPAC), William A. Rutala, PhD., M.PH., and David J.
Weber, M.D., M.PH. Centers for Disease Control and Pre-
vention, 2008. Guideline for Disinfection and Sterilization in
Healthcare Facilities. Atlanta, GA
Summary of Contents for CoaguChek XS Pro
Page 1: ...CoaguChek XS Pro P O I N T O F C A R E T E S T I N G Operator s Manual ...
Page 3: ...0 6950540001 02 2015 12 EN CAN Operator s Manual Version 5 0 CoaguChek XS Pro ...
Page 8: ...8 This page intentionally left blank ...
Page 21: ...The CoaguChek XS Pro Meter 2 21 2 The CoaguChek XS Pro Meter A B C D G I J F H K L E ...
Page 110: ...7 Review Results 110 This page intentionally left blank ...
Page 134: ...9 Maintenance and Care 134 This page intentionally left blank ...
Page 138: ...10 Troubleshooting 138 This page intentionally left blank ...
Page 146: ...A Appendix 146 Codabar Code 39 Code 93 Code 128 EAN 13 Interleaved 2 5 ...
Page 150: ...B Appendix 150 This page intentionally left blank ...
Page 154: ...154 This page intentionally left blank ...
Page 155: ......